1. Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors

    Evaluation of the optical performance for aspheric intraocular lenses in relation with tilt and decenter errors

    Purpose To evaluate and compare the effect of misalignment and tilt on the optical performance of different aspheric intraocular lens (IOL) designs. Methods Three aspheric IOLs with a different quantity of spherical aberration (SA) have been designed and the effect of IOL misalignment and tilt on the imaging quality of an eye model has been numerically assessed using a commercial optical design software. The prototypes have been manufactured by lathe turning and tested in vitro using the same optical bench (PMTF, Lambda-X) that complies with International Organization for Standardization standard 11979–2 requirements. Image quality was evaluated from the modulation ...