1. Uniform focusing with an extended depth range and increased working distance for optical coherence tomography by an ultrathin monolith fiber probe

    Uniform focusing with an extended depth range and increased working distance for optical coherence tomography by an ultrathin monolith fiber probe

    It is difficult to maintain high transverse resolution over an increased depth range using miniature probes for optical coherence tomography (OCT) due to the rapid divergence of light and the space limitation. To solve this problem, we introduce a fiber-based filter in the proposed probe to manipulate its output beam. Significant mode interference (MI) is exploited to enhance the depth of focus (DOF), and the mode phase difference is tuned to achieve a uniform axial intensity within the DOF. The magnified MI field instead of the diffracted one is adopted as the final pupil filter in the probe to increase ...