    Comparison of anterior segment measurements obtained using a swept-source optical coherence tomography biometer and a Scheimpflug–Placido tomographer

    Purpose To assess the repeatability and agreement of anterior segment measurements between an optical biometer (IOLMaster 700) using swept-source optical coherence tomography and a Scheimpflug–Placido tomographer (Sirius) in patients scheduled for corneal refractive surgery. Setting Eye Hospital of Wenzhou Medical University, Zhejiang, China. Design Prospective case series. Methods The parameters, including the central corneal thickness (CCT), anterior chamber depth (ACD), aqueous depth, keratometry (K) values, astigmatism, J0 (Jackson cross-cylinder with axes at 180 degree and 90 degrees) and J45 vectors (Jackson cross-cylinder with axes at 45 degrees and 135 degrees), and horizontal corneal diameter, were measured 3 times per ...