1. Predictive accuracy of partial coherence interferometry and swept-source optical coherence tomography for intraocular lens power calculation

    Predictive accuracy of partial coherence interferometry and swept-source optical coherence tomography for intraocular lens power calculation

    The purpose of this study is to compare the predictive accuracy of intraocular lens (IOL) calculations made with partial coherence interferometry (PCI, IOLMaster, version 5) and swept-source optical coherence tomography (SS-OCT, Argos). Axial length (AL), mean keratometry value (K), and anterior chamber depth (ACD) were obtained using PCI and SS-OCT optical biometers. Intraocular lens (IOL) power calculations were made using the Barret-Universal II, Haigis, Hoffer Q, SRK/T, and T2 formulas and compared the predictive accuracy between biometers. In 153 eyes (153 patients), axial length measurements made with PCI (24.65 ± 2.35 mm) and SS-OCT (24.62 ± 2.29 ...