1. Microscope-mounted intraoperative optical coherence tomography with actively tunable focus and optical path length

    Microscope-mounted intraoperative optical coherence tomography with actively tunable focus and optical path length

    We present a method for an intraoperative microscope-mounted Fourier-domain optical coherence tomography (FD-OCT) system to maintain high image contrast while dynamic adjusting focal planes. Because two imaging system with different imaging depth are integrated into one system, active control of OCT imaging conditions is indispensible for functioning high quality imaging modality. For the purpose of active adjustment of the focal plane, an electrically focus tunable lens (FTL) was used in the sample arm of the OCT system. Because the OCT image contrast at a depth is given by roll-off characteristics of the FD-OCT that is a function of difference in ...