1. Comparison of anterior segment optical coherence tomography angiography systems for corneal vascularisation

    Comparison of anterior segment optical coherence tomography angiography systems for corneal vascularisation

    Aim To newly describe a spectral-domain (SD) optical coherence tomography angiography (OCTA) for the cornea and directly compare two OCTA system scans of the same eyes with corneal vascularisation. Methods Cross-sectional, observational, comparative case series. We performed sequential OCTA scans (10 eyes of 10 subjects with corneal vascularisation,4 scans each eye) repeated using split-spectrum amplitude decorrelation algorithm angiography system (SSADA, AngioVue; Optovue Inc, USA) and SD OCTA (Angioscan; Nidek Co. Ltd, Japan) in the same region of interest. We analysed all scan images for repeatability, image quality and vessel density measurements and compared OCTA systems. Results We obtained substantial ...