Two-year vascular responses to drug-eluting stents with biodegradable polymer versus durable polymer: An optical coherence tomography sub-study of the NEXT

Background This study aimed to compare very late vascular response after stent implantation between everolimus-eluting stent (EES) with a thin, non-adhesive, durable, biocompatible fluorinated polymer and biolimus-eluting stent (BES) with a biodegradable polymer by optical coherence tomography (OCT). Methods and results In the NOBORI-BES Versus XIENCE V/PROMUS-EES Trial (NEXT), a formal OCT substudy investigated 48 patients (27 EES-treated lesions in 23 patients and 28 BES-treated lesions in 25 patients) with 2-year (18–30 months) follow-up imaging at 18 centers. The percentage of uncovered strut by neointima was significantly lower in EES compared with BES (2.1 ± 4.7% vs ...