Optimal threshold of postintervention minimum stent area to predict in-stent restenosis in small coronary arteries: An optical coherence tomography analysis

Objectives The aim of this study was to determine the best threshold of postintervention minimum stent area (MSA) assessed by optical coherence tomography (OCT) to predict long-term in-stent restenosis (ISR) for 2.5 mm-diameter everolimus-eluting stents (EES). Background Percutaneous coronary intervention (PCI) for small coronary arteries remains challenging. Stent underexpansion is a strong predictor of late ISR. Methods We performed a retrospective analysis of 69 lesions in 69 patients undergoing PCI with 2.5 mm-diameter stents using OCT for the assessment of postintervention MSA and subsequent 9-month angiographic follow-up. Results The rates of angiographic ISR and target lesion revascularization were ...