1. Scaffold and Edge Vascular Response Following Implantation of Everolimus-Eluting Bioresorbable Vascular Scaffold : A 3-Year Serial Optical Coherence Tomography Study

    Scaffold and Edge Vascular Response Following Implantation of Everolimus-Eluting Bioresorbable Vascular Scaffold : A 3-Year Serial Optical Coherence Tomography Study

    Objectives This study sought to investigate the in-scaffold vascular response (SVR) and edge vascular response (EVR) after implantation of an everolimus-eluting bioresorbable scaffold (BRS) using serial optical coherence tomography (OCT) imaging. Background Although studies using intravascular ultrasound have evaluated the EVR in metal stents and BRSs, there is a lack of OCT-based SVR and EVR assessment after BRS implantation. Methods In the ABSORB Cohort B (ABSORB Clinical Investigation, Cohort B) study, 23 patients (23 lesions) in Cohort B1 and 17 patients (18 lesions) in Cohort B2 underwent truly serial OCT examinations at 3 different time points (Cohort B1: post-procedure, 6 ...