1. Improved in vivo imaging of human blood circulation in the chorioretinal complex using phase variance method with new phase stabilized 1 μm swept-source optical coherence tomography (pv-SSOCT)

    Improved in vivo imaging of human blood circulation in the chorioretinal complex using phase variance method with new phase stabilized 1 μm swept-source optical coherence tomography (pv-SSOCT)

    We demonstrate the feasibility of our newly developed phase stabilized high-speed (100 kHz A-scans/s) 1 μm sweptsource optical coherence tomography (SSOCT) system with the phase-variance based motion contrast method for visualization of human chorioretinal complex microcirculation. Compared to our previously reported spectral domain (spectrometer based) phase-variance (pv)-SDOCT system it has advantages of higher sensitivity, reduced fringe washout for high blood flow speeds and deeper penetration in choroid. High phase stability SSOCT imaging was achieved by using a computationally efficient phase stabilization approach. This process does not require additional calibration hardware and complex numerical procedures. Our phase stabilization method ...