1. Comparison of Anterior Segment Optical Tomography Parameters Measured Using a Semi-Automatic Software to Standard Clinical Instruments

    Comparison of Anterior Segment Optical Tomography Parameters Measured Using a Semi-Automatic Software to Standard Clinical Instruments

    by Marcus Ang, Wesley Chong, Huiqi Huang, Wan Ting Tay, Tien Yin Wong, Ming-Guang He, Tin Aung, Jodhbir S. Mehta Objective To compare anterior segment parameters measured using a semi-automatic software (Zhongshan Angle Assessment Program, ZAP) applied to anterior segment optical coherence tomography (AS-OCT) images, with commonly used instruments. Methods Cross-sectional study of a total of 1069 subjects (1069 eyes) from three population-based studies of adults aged 40–80 years. All subjects underwent AS-OCT imaging and ZAP software was applied to determine anterior chamber depth (ACD), central corneal thickness (CCT), anterior and keratometry (K) – readings. These were compared to auto-refraction ...