Errors in neuroretinal rim measurement by Cirrus high-definition optical coherence tomography in myopic eyes
Background/aims To investigate the prevalence of, and factors associated with, errors in neuroretinal rim measurement by Cirrus high-definition (HD) spectral-domain optical coherence tomography (OCT) in myopic eyes. Methods Neuroretinal rim thicknesses of 255 myopic eyes were measured by Cirrus HD-OCT. The prevalence of, and factors associated with, optic disc margin detection error and cup margin detection error were assessed by analysing 72 cross-sectional optic nerve head (ONH) images obtained at 5° intervals for each eye. Results Among the 255 eyes, 45 (17.6%) had neuroretinal rim measurement errors; 29 (11.4%) had optic disc margin detection errors at the ...

Login to comment.