1. Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomography

    Repeatability and reproducibility of anterior ocular biometric measurements with 2-dimensional and 3-dimensional optical coherence tomography
    Purpose To evaluate the repeatability and reproducibility of central corneal thickness (CCT), anterior chamber depth (ACD), and anterior chamber width (ACW) measurements using 3-dimensional (3-D) corneal and anterior segment optical coherence tomography (CAS-OCT) and 2-dimensional (2-D) anterior segment OCT (AS-OCT). Setting Department of Ophthalmology, Institute of Clinical Medicine, University of Tsukuba, Ibaraki, Japan. Design Nonrandomized clinical trial. Methods The CCT, ACD, and ACW were measured in normal eyes using a prototype 3-D swept-source CAS-OCT device and a 2-D time-domain AS-OCT device (Visante). The coefficient of repeatability and reproducibility and the intraclass correlation coefficient (ICC) were calculated to evaluate the repeatability ...
    Read Full Article