1. Automated Segmentation of Neural Canal Opening and Optic Cup in 3-D Spectral Optical Coherence Tomography Volumes of the Optic Nerve Head

    Automated Segmentation of Neural Canal Opening and Optic Cup in 3-D Spectral Optical Coherence Tomography Volumes of the Optic Nerve Head
    Purpose: To develop an automated approach for segmenting the neural canal opening (NCO) and cup at the level of retinal pigment epithelium (RPE)/Bruch's membrane (BM) complex in spectral-domain optical coherence tomography (SD-OCT) volumes. To investigate the correspondence and discrepancy of the NCO-based metrics and clinical disc margin on fundus photographs in glaucoma subjects. Methods: SD-OCT scans (CirrusTM HD-OCT) and corresponding stereo fundus photographs (Nidek 3Dx) of the optic nerve head (ONH) are obtained from 68 eyes of 34 patients with glaucoma or glaucoma suspicion. Manual planimetry is performed by three glaucoma experts to delineate a reference standard (RS ...