1. Polarization Sensitive Subcutaneous and Muscular Imaging Based on Common Path Optical Coherence Tomography Using Near Infrared Source

    In this paper, we describe a polarization sensitive (PS) subcutaneous and muscular imaging system based on common path optical coherence tomography (CP-OCT) using a near infrared source. The axial and lateral resolutions of the PS-OCT system are 9 and 6 μm, respectively. The main goal of this work is to build a high-resolution and minimally invasive optical imager for examining various kinds of cutaneous substructures with intrinsic or form birefringence. The internal structural information is extracted by the real-time signal analysis (Fourier Transform) of the modulated spectral intensity depending on the beam and tissue birefringence. The preliminary results using fresh ...