1. Variability of optical coherence tomography in assessment of optic disc and retinal nervous fibre layer

    (Article in Romanian) PURPOSE: To study the short-term variability of retinal nerve fiber layer (RNFL) and optic disc measurements using Stratus OCT. METHOD: Transversal comparative study with repeated measurements. For each eye, the Fast Optic Disc and Fast RNFL protocols were repeated twice in the same session. The resulted measurements were compared using correlation and variation coefficients. RESULTS: 62 eyes (34 patients) were included in the study. For every analyzed parameter there was a highly significant correlation between the first and the second measurement (p < 0.001). For the C/D ratio (vertical, horizontal and area) the correlation coefficients varied ...