Comparison of acquisition rate and agreement of axial length with two swept-source optical coherence tomographers and a partial coherence interferometer

Purpose To assess the acquisition rate and agreement of axial length (AL) measurements with the OA-2000, Anterion, and IOLMaster 500 in cataractous patients. Methods In total, 298 eyes of 191 cataractous patients were enrolled and scanned with the three devices in random order. The success rate of AL measurements per device was calculated and a chi-square test was utilized to identify the differences in acquisition rate between the three devices. Logistic regression analysis was applied to evaluate the association of different cataract types and severity with the AL measurement acquisition rate. Bland–Altman plots were mapped to appraise the agreement ...