1. Agreement of intraocular lens power calculation between 2 SS-OCT–based biometers

    Agreement of intraocular lens power calculation between 2 SS-OCT–based biometers

    Purpose: To assess agreement of measurements by 2 swept-source optical coherence tomography biometers and to evaluate the prediction error (PE) in intraocular lens power calculation with 7 formulas. Setting: Tertiary public eye hospital. Design: Consecutive observational. Methods: Axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness (LT), and corneal diameter (CD) were measured with the IOLMaster 700 (Biometer A) and Anterion (Biometer B). Agreement was quantified by the limits of agreement and concordance correlation coefficient (CCC). The PE, the median absolute error, and the mean absolute error of the Barrett Universal II, EVO 2.0, Haigis, Hoffer Q ...