1. Serious complications of cosmetic NewColorIris implantation

    Purpose: This case report describes serious postoperative complications and markedly elevated intraocular pressure (IOP) associated with the NewColorIris cosmetic implant. Methods: We report an interventional case series of two patients who suffered multiple complications after NewColorIris implantation carried out in Panama. Assessment included visual acuity, photography, endothelial cell count and anterior segment optical coherence tomography (OCT) when possible. Results: Both patients presented with endothelial cell loss, uveitis, pigment dispersion and elevated IOP. Anterior segment OCT demonstrated irregularities in the position and configuration of the implants within the anterior chamber with resultant areas of implant[ndash]iris and implant[ndash]endothelial ...
    Read Full Article

    Login to comment.

  1. Categories

    1. Applications:

      Art, Cardiology, Dentistry, Dermatology, Developmental Biology, Gastroenterology, Gynecology, Microscopy, NDE/NDT, Neurology, Oncology, Ophthalmology, Other Non-Medical, Otolaryngology, Pulmonology, Urology
    2. Business News:

      Acquisition, Clinical Trials, Funding, Other Business News, Partnership, Patents
    3. Technology:

      Broadband Sources, Probes, Tunable Sources
    4. Miscellaneous:

      Jobs & Studentships, Student Theses, Textbooks
  2. Topics Mentioned

  3. Authors