1. The Precision of a New Anterior Segment Optical Coherence Tomographer and Its Comparison With a Swept-Source OCT-Based Optical Biometer in Patients With Cataract

    The Precision of a New Anterior Segment Optical Coherence Tomographer and Its Comparison With a Swept-Source OCT-Based Optical Biometer in Patients With Cataract

    PURPOSE: To assess the precision of a new spectral-domain optical coherence tomographer (SD-OCT)/Placido topographer (MS-39; CSO) and its comparison with a swept-source OCT (SS-OCT) biometer (Argos; Movu, Inc) in patients with cataract. METHODS: Fifty-three right eyes from 53 patients were examined by two experienced operators three times using both devices randomly. Employing the within-subject standard deviation (S w ), test–retest variability, coefficient of variation, and intraclass correlation coefficient to evaluate intraoperator repeatability and interoperator reproducibility; the double-angle plots to analyze astigmatism; and Bland-Altman plots and 95% limits of agreement to verify the agreement between devices. RESULTS: The SD-OCT/Placido ...