    Agreement between two swept-source OCT biometers and a Scheimpflug partial coherence interferometer

    Purpose: To evaluate the agreement between different parameters obtained with two swept-source optical coherence tomography (SS-OCT)-based biometers and one Scheimpflug camera with partial coherence interferometry (PCI). Setting: Oftalvist, Spain. Design: Prospective case series. Methods: Biometry was performed in 49 eyes using 3 optical biometers: ANTERION SS-OCT, IOLMaster 700 SS-OCT, and Pentacam AXL PCI. Keratometry (K), J0 and J45 vectors, anterior chamber depth (ACD), central corneal thickness (CCT), white-to-white (WTW), lens thickness (LT), and axial length (AL) were measured with each device. Bland-Altman analysis was applied. Results: There were no statistically significant differences for K1, K2, J0 and J45 between ...