1. Twelve-month clinical and imaging outcomes of the uncaging DynamX Bioadaptor System

    Aims- We assessed the safety and efficacy of the DynamX™ Novolimus-Eluting Coronary Bioadaptor System, a novel device that initially acts as a second-generation drug-eluting stent, but after six months frees the vessel through uncaging elements. Methods and results- This multi-center study enrolled 50 patients with single de novo lesions. In-device acute lumen gain was 1.61±0.34 mm, and device and procedure success was 100%. Through 12 months, two target lesion failures occurred, both were cardiac deaths (day 255 and 267 post-procedure). No definite or probable device thrombosis was observed. Mean late lumen loss was 0.12±0.18 ...

    Read Full Article